Category Products
Partner Links
Material test chambers | FP 720
Untuk memesan, silahkan tekan tombol ORDER. Anda bisa memesan barang secara Online.
Anda juga bisa memesan via telephone. Silakan hubungi sales consultant, Hotline kami di, sales@alatlabor.com
sales Consultant kami akan menjelaskan detail produk dan harga terbaru. atau email ke alamatAbout Material test chambers | FP 720
Material test chamber with mechanical convection
A BINDER material test chamber with mechanical convection of the FP series provides reliably short drying times and particularly fast heating – even for chambers under full loads.
Features :
Features | Customer benefits | Characteristics |
APT.line™ temperature technology |
|
APT.line for maximum precision ™
|
Performance characteristics |
|
|
Inner chamber concept |
|
|
Quality |
|
|
Accessories and Services |
|
Comprehensive product portfolio
|
Spesification :
FP 720 | |
Exterior dimensions | |
Width (mm) | 1235 |
Height (incl. casters) (mm) | 1530 |
Depth (mm) | 865 |
Plus door handle and connection (mm) | 105 |
Wall clearance, rear (mm) | 100 |
Wall clearance, side (mm) | 160 |
Exhaust duct (outer Ø mm) | 52 |
Steam space volume (l) | 869 |
Number of doors (ea.) | 2 |
Interior dimensions | |
Width (mm) | 1000 |
Height (mm) | 1200 |
Depth (mm) | 600 |
Interior volume (l) | 720 |
Racks (number standard/max.) | 2 / 15 |
Load per rack (kg) | 45 |
Permitted total load (kg) | 120 |
Weight (empty) (kg) | 184 |
Temperature data | |
Temperature range approx. 5 °C above ambient temperature to (°C) | 300 |
Temperature variation | |
at 70 °C (± K) | 1 |
at 150 °C (± K) | 2 |
at 300 °C (± K) | 5,5 |
Temperature fluctuation (± K) | 0,3 |
Warm-up time 1) | |
to 70 °C (min.) | 25 |
to 150 °C (min.) | 39 |
to 250 °C (min.) | 65 |
Recovery time after doors were open for 30 sec. 1) | |
at 70 °C (min.) | 2 |
at 150 °C (min.) | 20 |
at 300 °C (min.) | 24 |
Ventilation data | |
Ventilation | |
at 70 °C (x/h) | 11 |
at 150 °C (x/h) | 12 |
at 300 °C (x/h) | 10 |
Electrical data | |
IP protection class acc. to EN 60529 | IP 20 |
Voltage (± 10%) 50 / 60 Hz (V) | 400 3N~ |
Nominal power (kW) | 5 |
Energy consumption | |
at 70 °C (W) | 570 |
at 150 °C (W) | 1320 |
at 300 °C (W) | 2600 |